When is a premarket notification (510(k)) submission required before marketing a medical device under 21 CFR 807.81? Answer in under 100 words.
Details / How it works
Under 21 CFR 807.81, each person required to register under §807.20 must submit a premarket notification (510(k)) to FDA at least 90 days before introducing a device into interstate commerce for commercial distribution. The requirement applies to:
- First-of-a-kind devices – those not substantially equivalent to a device in commercial distribution before May 28, 1976, or to a device introduced after that date that has since been reclassified into Class I or II.
- First-time introduction by a registrant – even if the device itself has been marketed by others.
- Significant changes or modifications – changes that could significantly affect safety or effectiveness (e.g., design, material, chemical composition, energy source, manufacturing process) or a major change in intended use.
Numbers & thresholds
| Factor | Threshold |
|---|---|
| Advance notice | At least 90 days before commercial distribution |
| Predicate date | Devices in commercial distribution before May 28, 1976 are the baseline for substantial equivalence |
| Reclassification | Devices introduced after May 28, 1976 and later reclassified into Class I or II are also predicates |
| Significant change | A modification that could significantly affect safety or effectiveness triggers a new submission |
| Major intended use change | A major change in intended use triggers a new submission |
Exceptions & edge cases
A premarket notification is not required when:
- A Premarket Approval Application (PMA) under section 515 of the Act is pending before FDA.
- A petition to reclassify under section 513(f)(2) is pending.
- The change is consistent with a predetermined change control plan (PCCP) cleared under section 515C of the Act.
Additionally, the appropriate FDA Center Director may determine that a written request for an exception or alternative under §801.128 or §809.11 satisfies the premarket notification requirement for significant changes under paragraph (a)(3). Manufacturers of radiation-emitting electronic products must also comply with part 1002 reporting requirements.
Sources
- [1]U.S. National Archives, Electronic Code of Federal Regulations — 21 CFR 807.81 — When a Premarket Notification (510(k)) Submission Is Required · as of 2026-08-26