Minds → FDA Regulatory → What changes to a marketed medical device require a new premarket notification? Answer in under 100 words. Cite only eCFR (www.ecfr.gov) sources.
US Federal

What changes to a marketed medical device require a new premarket notification? Answer in under 100 words. Cite only eCFR (www.ecfr.gov) sources.

Under 21 CFR 807.81(a)(3) (as amended at 89 FR 18792, Mar. 15, 2024), a new premarket notification (510(k)) is required when a device already in commercial distribution is about to be significantly changed or modified in design, components, method of manufacture, or intended use. The rule defines two categories of significant change: (i) any change that could significantly affect safety or effectiveness (e.g., in design, material, chemical composition, energy source, or manufacturing process), and (ii) a major change or modification in the intended use of the device.
Last verified: 2026-08-27 · Sources checked: 1/1 resolving · Next scheduled review: 2026-11-25

Details / How it works

Per 21 CFR 807.81(a)(3), a person who is required to register under § 807.20 must submit a premarket notification at least 90 days before introducing a significantly changed or modified device into interstate commerce.

The regulation identifies two categories of significant change or modification that trigger a new 510(k):

  • Changes affecting safety or effectiveness — e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process.
  • Major change in intended use of the device.

The baseline list of devices exempt from notification (because they are substantially equivalent) is set out in § 807.81(a)(1), referencing devices in commercial distribution before May 28, 1976, and devices reclassified into class I or II after that date.

Numbers & thresholds

Item Value Source
Minimum lead time before commercial distribution 90 days 21 CFR 807.81(a)
Pre-1976 commercial distribution baseline date May 28, 1976 21 CFR 807.81(a)(1)(i)
Most recent amendment date Mar. 15, 2024 (89 FR 18792) 21 CFR 807.81

Exceptions & edge cases

Under 21 CFR 807.81(b)(1), a premarket notification is not required when:

  • A premarket approval application (PMA) under section 515 of the FD&C Act, or a petition to reclassify under section 513(f)(2), is pending before FDA; or
  • There is a Predetermined Change Control Plan (PCCP) cleared under section 515C of the FD&C Act, provided the change is consistent with that PCCP.

Under 21 CFR 807.81(b)(2), the appropriate FDA Center Director may also determine that a written request for an exception or alternative granted under § 801.128 or § 809.11 satisfies the § 807.81(a)(3) requirement.

Under 21 CFR 807.81(c), owners or operators that manufacture radiation-emitting electronic products (as defined in § 1000.3) must additionally comply with the reporting requirements of part 1002.

Sources

  1. [1]U.S. National Archives, Electronic Code of Federal Regulations — 21 CFR 807.81 — When a Premarket Notification (510(k)) Submission Is Required · as of 2026-08-27

Ask this MIND

This answer is static and source-locked. Ask FDA Regulatory your version of this question.

Related questions

This is general information, not advice. FDA Regulatory summarizes primary sources (FDA, HHS) as of the dates shown. It is not tax/legal/immigration advice and doesn’t account for your situation. Consult a qualified professional before acting. About this MIND