Per 21 CFR 807.81(a)(3), a person who is required to register under § 807.20 must submit a premarket notification at least 90 days before introducing a significantly changed or modified device into interstate commerce.
The regulation identifies two categories of significant change or modification that trigger a new 510(k):
- Changes affecting safety or effectiveness — e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process.
- Major change in intended use of the device.
The baseline list of devices exempt from notification (because they are substantially equivalent) is set out in § 807.81(a)(1), referencing devices in commercial distribution before May 28, 1976, and devices reclassified into class I or II after that date.
Numbers & thresholds
| Item | Value | Source |
|---|---|---|
| Minimum lead time before commercial distribution | 90 days | 21 CFR 807.81(a) |
| Pre-1976 commercial distribution baseline date | May 28, 1976 | 21 CFR 807.81(a)(1)(i) |
| Most recent amendment date | Mar. 15, 2024 (89 FR 18792) | 21 CFR 807.81 |
Exceptions & edge cases
Under 21 CFR 807.81(b)(1), a premarket notification is not required when:
- A premarket approval application (PMA) under section 515 of the FD&C Act, or a petition to reclassify under section 513(f)(2), is pending before FDA; or
- There is a Predetermined Change Control Plan (PCCP) cleared under section 515C of the FD&C Act, provided the change is consistent with that PCCP.
Under 21 CFR 807.81(b)(2), the appropriate FDA Center Director may also determine that a written request for an exception or alternative granted under § 801.128 or § 809.11 satisfies the § 807.81(a)(3) requirement.
Under 21 CFR 807.81(c), owners or operators that manufacture radiation-emitting electronic products (as defined in § 1000.3) must additionally comply with the reporting requirements of part 1002.
Sources
- [1]U.S. National Archives, Electronic Code of Federal Regulations — 21 CFR 807.81 — When a Premarket Notification (510(k)) Submission Is Required · as of 2026-08-27
Ask this MIND
This answer is static and source-locked. Ask FDA Regulatory your version of this question.
Related questions
This page is one mind’s knowledge, published.
MIND turns your documents into a queryable, citable knowledge graph.
Make your own MIND →